EUROPA B EVO AUTOCLAVE - 15 litres - 230V
• External size: 510 x 390 x 590 mm (W x H x D)
• Chamber size: 245 x 320 mm (Ø x dept)
• Operating voltage 230V - 50/60 Hz
• Max absorbed power: 1,700 W
• Weight: 54 kg
• Built in printer: YES
• Manual closing: YES
• Fitting for pura deioniser: YES
• PID system: YES
• Cycles: 121°C Wrapped
• Working pressure min-max: 1.04 - 1.30
• Minimum time of the total cycle*: 60 minutes (25+18+17)
• Cycles: 121°C Unwrapped
• Working pressure min-max: 1.04 - 1.30
• Minimum time of the total cycle*: 49 minutes (25+18+6)
• Cycles: 134°C Flash
• Working pressure min-max: 2.02 - 2.38
• Minimum time of the total cycle*: 31 minutes (25+4+2)
• Cycles: 134°C Wrapped
• Working pressure min-max: 2.02 - 2.38
• Minimum time of the total cycle*: 47 minutes (25+5+17)
• Cycles: 134°C Unwrapped
• Working pressure min-max: 2.02 - 2.38
• Minimum time of the total cycle*: 36 minutes (25+5+6)
• Cycles: 134°C Prion
• Working pressure min-max: 2.02 - 2.38
• Minimum time of the total cycle*: 62 minutes (25+20+17) *The total time includes: vacuum, pre-heating time, cycle and drying phase. The times are to be intended for a fully loaded autoclave. All sterilization cycles have 3 stages of vacuum. Vacuum and pre-heating time can vary depending on the conditions of autoclave from 25 to 35 minutes.
TECNICAL SPECIFICATIONS
• External size: 510 x 390 x 590 mm (W x H x D)
• Chamber size: 245 x 320 mm (Ø x dept)
• Operating voltage 230V - 50/60 Hz
• Max absorbed power: 1,700 W
• Weight: 54 kg
• Built in printer: YES
• Manual closing: YES
• Fitting for pura deioniser: YES
• PID system: YES
• Cycles: 121°C Wrapped
• Working pressure min-max: 1.04 - 1.30
• Minimum time of the total cycle*: 60 minutes (25+18+17)
• Cycles: 121°C Unwrapped
• Working pressure min-max: 1.04 - 1.30
• Minimum time of the total cycle*: 49 minutes (25+18+6)
• Cycles: 134°C Flash
• Working pressure min-max: 2.02 - 2.38
• Minimum time of the total cycle*: 31 minutes (25+4+2)
• Cycles: 134°C Wrapped
• Working pressure min-max: 2.02 - 2.38
• Minimum time of the total cycle*: 47 minutes (25+5+17)
• Cycles: 134°C Unwrapped
• Working pressure min-max: 2.02 - 2.38
• Minimum time of the total cycle*: 36 minutes (25+5+6)
• Cycles: 134°C Prion
• Working pressure min-max: 2.02 - 2.38
• Minimum time of the total cycle*: 62 minutes (25+20+17) *The total time includes: vacuum, pre-heating time, cycle and drying phase. The times are to be intended for a fully loaded autoclave. All sterilization cycles have 3 stages of vacuum. Vacuum and pre-heating time can vary depending on the conditions of autoclave from 25 to 35 minutes.
- CND
- Z12011305
- GMDN
- 42162
- Classe Medicale
- II B
- Dispositivo Medico
- Dispositivo medico
- unità
- 1 pz.
- NSIS
- 1570217