MT 10 HANDHELD TYMPANOMETER WITH IPSI-REFLEX
• Analysis performed: Peak compliance; peak pressure, gradient; equivalent ear canal volume
• Probe tone: 226 Hz
• Pressure levels: +200 daPa to -400 daPa
• Ear volume measurement range: 0.2 ml to 5 ml
• Sweep speed: Typically 200-300 daPa/sec;depending on ear & cavity volume
• Pressure limits (safety shut off): +600 to -800 daPa Reflex measuremets
• Measurement mode: ipsilateral
• Reflex tone levels: 500 Hz, 1 kHz, 2 kHz, 4 kHz
• Four reflex levels: 100 dB with 5 dB or 10 dB steps; 95 dB, 90 dB or 85 dB with 5 dB steps
• Reflex analysis: - Reflex pass/refer at each level tested; - Maximum amplitude of each reflex (seen on printed report and PC report); - Pressure at which reflex was performed Data management
• Data storage: 30 records. Any recording can be stored once the tympanogram is viewed
• Data held: Patient Initials, Tympanogram and Reflex graphs and analysis for Left Ear and/or Right Ear, Time and Date of recording, which ears were tested, whether or not the record has been printed and for sent to a PC, parameters printed and/or sent to a PC, parameters used for analysis Other data
• Internal backup power supply: >30 days without main batteries fitted
• Battery Lifetime: Approx. 400 tests (Alkaline AA batteries)
• Supported printer: Martel MCP8830 or Sanibel MPTII
• Interface: Infra-red, IrDA hardware, 9600 baud
• Information printed: - Space for patient and clinician's details - Tympanogram analysis parameters - Tympanogram, Reflex analysis parameters - Reflex graph, Serial Number of device - Last and Next Due Calibration dates
• Standards: Safety: IEC 60601-1. EMC: IEC 60601-1-2
• Impedance standard: lEe 60645-5, Type 2 Tympanometer
TECNICAL SPECIFICATIONS
Tympanometry
• Analysis performed: Peak compliance; peak pressure, gradient; equivalent ear canal volume
• Probe tone: 226 Hz
• Pressure levels: +200 daPa to -400 daPa
• Ear volume measurement range: 0.2 ml to 5 ml
• Sweep speed: Typically 200-300 daPa/sec;depending on ear & cavity volume
• Pressure limits (safety shut off): +600 to -800 daPa Reflex measuremets
• Measurement mode: ipsilateral
• Reflex tone levels: 500 Hz, 1 kHz, 2 kHz, 4 kHz
• Four reflex levels: 100 dB with 5 dB or 10 dB steps; 95 dB, 90 dB or 85 dB with 5 dB steps
• Reflex analysis: - Reflex pass/refer at each level tested; - Maximum amplitude of each reflex (seen on printed report and PC report); - Pressure at which reflex was performed Data management
• Data storage: 30 records. Any recording can be stored once the tympanogram is viewed
• Data held: Patient Initials, Tympanogram and Reflex graphs and analysis for Left Ear and/or Right Ear, Time and Date of recording, which ears were tested, whether or not the record has been printed and for sent to a PC, parameters printed and/or sent to a PC, parameters used for analysis Other data
• Internal backup power supply: >30 days without main batteries fitted
• Battery Lifetime: Approx. 400 tests (Alkaline AA batteries)
• Supported printer: Martel MCP8830 or Sanibel MPTII
• Interface: Infra-red, IrDA hardware, 9600 baud
• Information printed: - Space for patient and clinician's details - Tympanogram analysis parameters - Tympanogram, Reflex analysis parameters - Reflex graph, Serial Number of device - Last and Next Due Calibration dates
• Standards: Safety: IEC 60601-1. EMC: IEC 60601-1-2
• Impedance standard: lEe 60645-5, Type 2 Tympanometer
- Classe Medicale
- II A
- Dispositivo Medico
- Dispositivo medico
- unità
- 1 pz.